Robot-Assisted Deep Brain Stimulation for Parkinson’s Disease : Accuracy, Workflow, Clinical Outcomes, and Safety
Article information
Abstract
Robot-assisted deep brain stimulation (DBS) has emerged as a major advancement in stereotactic and functional neurosurgery, enhancing precision, reproducibility, and efficiency in targeting deep brain structures for Parkinson’s disease. This comprehensive review synthesizes the current state of robotic DBS platforms, comparing their technical mechanisms, targeting accuracy, clinical outcomes, workflow efficiency, and safety profiles. Across diverse robotic systems, reported robotic targeting errors are typically around 1 mm, with values ranging from submillimeter to approximately 1.5 mm. Clinical outcomes, including 40–60% improvements in Unified Parkinson’s disease rating scale (UPDRS)-III scores and significant reductions in dopaminergic medication, parallel those achieved with traditional approaches, while complication rates remain within the expected range. Robotic systems further enhance procedural efficiency through automated trajectory alignment, image-guided verification, and reduced intraoperative variability, with a clear learning-curve effect observed across centers. This review summarizes published evidence, highlights the unique strengths of robot assisted DBS, and discusses the evolving role of robotic DBS in modern neurosurgical practice.
INTRODUCTION
Parkinson’s disease (PD) is a progressive neurodegenerative disorder marked by tremor, rigidity, and bradykinesia [33]. While pharmacologic therapy remains the initial treatment, patients with advanced-stage PD who experience refractory motor symptoms or fluctuations often benefit substantially from deep brain stimulation (DBS) [6,20,38]. For such reasons, DBS is now an established therapy for advanced PD, particularly in patients with motor complications unresponsive to optimal medical treatment [30].
Precise electrode placement within the target nucleus–typically the subthalamic nucleus or the globus pallidus internus–is essential for maximizing therapeutic benefit and minimizing adverse effects [14]. Traditional DBS implantation relies on frame-based stereotaxy, intraoperative microelectrode recording (MER), and awake surgery for physiological confirmation. However, these conventional methods have limitations including patient discomfort, prolonged operative times, and potential human error during trajectory alignment [18].
Robotic assistance in DBS (robot-assisted DBS, R-DBS) has emerged as a promising innovation to address these challenges [32]. By enabling automation of trajectory alignment, frameless registration, and integration of intraoperative imaging-functions that vary among individual platforms–robotic systems aim to enhance reproducibility, streamline workflow, and reduce targeting errors. As robotic technologies continue to be incorporated into functional neurosurgical practice, an updated and clinically oriented synthesis of their role in PD surgery is warranted.
While several recent reviews and meta-analyses have broadly examined robotic stereotactic systems or R-DBS across mixed disease indications, fewer studies have specifically focused on R-DBS in PD. Given that PD represents the most common contemporary indication for DBS, a disease-specific synthesis may offer greater clinical relevance. Accordingly, this review aims to contextualize robotic assistance within PD surgery by emphasizing PD-specific clinical outcomes, operative workflows, and platform-level considerations relevant to real-world functional neurosurgical practice.
Scope of the review and literature search
This review was conducted as a structured narrative review of the published literature on R-DBS for PD. A literature search was performed using major electronic databases, including PubMed and Embase, covering studies published from database inception through August 2025. Search terms included combinations of “robot-assisted,” “robotic,” “deep brain stimulation,” “DBS,” and “Parkinson’s disease.” Peer-reviewed publications evaluating R-DBS specifically in patients with PD were considered for inclusion. Studies focusing on robotic stereotactic systems without application to DBS, or R-DBS performed for indications other than PD, were not included. Early case reports, first-in-human experiences, and technical notes were intentionally included to capture the historical development and technical evolution of robotic DBS platforms in PD. Given the narrative scope of this review, no formal risk-of-bias assessment was performed. We acknowledge that the existing robotic DBS literature predominantly reports favorable outcomes, and therefore the potential for publication bias should be considered when interpreting the findings summarized herein.
BASIC CONCEPT OF ROBOTIC SYSTEMS IN NEUROSURGERY
Robotic systems used in surgery can be classified according to their operating mechanism and degree of autonomy [16]. The most well-known configuration is the master-slave system, in which the surgeon remotely controls robotic arms in real time to perform the operation. This approach enhances visualization and allows precise instrument manipulation through small incisions or narrow anatomical spaces. The surgeon operates the robot while viewing a high-definition stereoscopic monitor, enabling minimally invasive access to regions that are otherwise difficult to reach using conventional techniques.
Another class of surgical robots operates semi-autonomously, following a preplanned trajectory or treatment protocol input by the operator. A representative example is the robotic radiosurgery system, which automatically moves and delivers focused radiation beams from multiple angles according to a predefined treatment plan. This design ensures accurate dose delivery to the target while reducing the time required for manual equipment adjustment.
Lastly, another configuration which is commonly used in stereotactic and functional neurosurgery, combines robotic precision with direct human execution. In these systems, the robot guides instruments such as electrodes or catheters toward the target defined on preoperative imaging, while the final insertion and manipulation are performed by the surgeon. Such systems enhance alignment accuracy and stability while maintaining human oversight during the critical procedural steps. This image-guided, surgeon-controlled robotic paradigm forms the basis of current applications in DBS, which is the primary focus of this review.
ACCURACY, REPRODUCIBILITY AND EFFICIENCY OF ROBOTIC SYSTEMS IN DBS
The key performance indicators of robotic systems in DBS are accuracy, reproducibility, and efficiency.
Accuracy is typically quantified by parameters such as radial error and Euclidian (vector) error, which describe the deviation between planned and actual electrode trajectories. Among these, radial error is considered the most clinically relevant, since the lead tip spans a sufficient depth range along the trajectory, allowing fine adjustment through MER without altering the trajectory itself.
Reproducibility reflects the consistency of targeting accuracy across cases. Although the mean or median radial error provides a primary measure of precision, the variance or range of error distribution is equally important, as smaller variability implies reduced risk of serious complications arising from outlier deviations.
Efficiency can be evaluated using multiple time-related parameters, including total procedural time (from surgical preparation to completion), anesthesia duration, surgical time, and robot setup time. While procedural efficiency contributes to operating-room workflow optimization, it should be emphasized that in DBS, the primary objective of robotic systems is to enhance accuracy and eliminate human-related sources of error; time reduction remains a secondary, albeit beneficial, goal.
CURRENTLY AVAILABLE STEREOTACTIC ROBOTIC SYSTEMS APPLIED TO DBS
The concept of robotic stereotaxy in neurosurgery dates back to Kwoh et al. [15], who developed the first robotic system capable of performing computed tomography-guided brain biopsies. This pioneering work demonstrated the feasibility of robotic precision in cranial procedures and laid the foundation for modern R-DBS.
In the following decades, robot-assisted stereotactic systems have evolved into practical alternatives to conventional frame-based and frameless techniques (Table 1). Platforms such as ROSA (Zimmer Biomet), Neuromate (Renishaw), and Renaissance (Mazor Robotics/Medtronic) are among the most extensively clinically reported in DBS. Among these, Neuromate holds the longest record of continuous clinical use, whereas ROSA has gained global adoption owing to its workflow flexibility and compatibility with both frame-based and frameless approaches.
Recently, several new robotic systems have emerged across different regions, expanding the technological diversity in DBS. In Korea, KYMERO (Koh Young Technology) developed a frame-based bed attachable robotic system integrating optical tracking and navigation and its upgraded version, Geniant, has recently obtained United States Food and Drug Administration clearance. In China, platforms such as SR1 (Sinovation) and REMEBOT (Remebot, Beijing) have demonstrated clinical feasibility and accuracy comparable to established Western systems. In Europe and North America, modular and compact robotic devices such as Cirq (Brainlab, Germany) and Stealth Autoguide (Medtronic, USA)—have been adapted for cranial procedures and possess potential applicability for DBS, although published data remain limited. These developments reflect the increasing global interest in robotic stereotaxy and the convergence of diverse engineering approaches to improve neurosurgical precision.
Surgical strategies also vary across centers. Some institutions continue to perform awake DBS with MER for physiological confirmation, whereas others have adopted “asleep” imageguided DBS under general anesthesia [1]. Most robotic systems can operate in frame-based or frameless configurations and integrate easily with intraoperative imaging for trajectory verification (Table 1). Reported targeting accuracy typically falls within the submillimeter to approximately 1 mm range, which is comparable to or superior to traditional frame-based stereotaxy. In addition to high targeting precision, robotic assistance provides workflow benefits such as automated trajectory alignment and elimination of manual coordinate conversion.
Each robotic platform offers unique technical strengths and workflow efficiencies; however, no single system has yet demonstrated clear superiority. Instead, these differences highlight varying institutional philosophies, surgical preferences, and regulatory environments.
ACCURACY OF R-DBS
Accurate electrode placement is fundamental to the success of DBS [2,14,39]. Robotic systems were developed to enhance targeting precision and reproducibility while minimizing operator-dependent variability inherent in manual stereotaxy. Multiple clinical studies have demonstrated that R-DBS achieves targeting errors within the submillimeter to 1.5 mm range, with reported accuracy generally comparable to that of conventional frame-based methods [3,4,7,8,10,11,13,17,19,21-24,26-28,34-36,40].
In large clinical series, VanSickle et al. [35] reported a mean radial error of 0.85±0.38 mm using the Renaissance system with intraoperative computed tomography (CT) verification in 128 patients, while Moran et al. [24] documented a mean radial error of 0.6±0.33 mm in 113 patients using the Neuromate system. Xu et al. [40] observed a mean error of 0.97±0.31 mm using the conventional registration method in the ROSA system, which improved to 0.65±0.23 mm after a modified registration method. Fontaine et al. [9] similarly reported a mean error of approximately 0.7 mm with Neuromate system in a frameless workflow.
A meta-analysis encompassing DBS and SEEG by Philipp et al. [29] indicated that frameless techniques were associated with a mean 0.50-mm (95% confidence interval [CI], 0.17–0.84) increase in error, while the use of intraoperative physiologic testing contributed a 0.45-mm (95% CI, 0.10–0.80) increase.
More recently Huang et al. [12] conducted a pooled meta-analysis focused on patients with PD, showing a mean accuracy of 1.09 mm (95% CI, 0.87–1.30), confirming that robotic approaches are non-inferior to frame-based stereotaxy. Dashtkoohi et al. [4] further reported a pooled mean radial error of 1.04 mm (95% CI, 0.93–1.15) and a euclidean error of 1.19 mm (95% CI, 0.94–1.45). Direct comparisons between R-DBS and conventional manual DBS showed no statistically significant difference in targeting precision (standardized mean difference, -0.56; 95% CI, -1.35 to 0.22; p=0.16) [4]. Likewise, Wan et al. [37] reported a similar trend toward improved accuracy with robotic guidance but overlapping confidence intervals and no increase in complication rates (odds ratio [OR], 1.79; 95% CI, 0.79–4.05; p=0.17). Collectively, these findings support the reproducibility and safety of robotic stereotaxy in DBS and indicate that R-DBS achieves targeting accuracy that is non-inferior to conventional frame-based techniques.
Despite this high level of precision, several studies have described directional bias in targeting error depending on the robotic registration or surgical technique. Mei et al. [23] noted a mild posterior drift in robotic trajectories, whereas traditional frame-based implants tended to deviate medially and posteriorly. Contributing factors likely include brain shift, pneumocephalus, cerebrospinal fluid loss, and registration inaccuracy. Jin et al. [13] compared awake versus asleep R-DBS and implemented intraoperative registration correction to minimize drift, reporting mean radial errors of 0.71±0.25 mm (general anesthesia) and 0.76±0.23 mm (local anesthesia).
Overall, R-DBS appears to reduce the variance of targeting error compared with manual methods, suggesting greater consistency across cases rather than absolute gains in mean accuracy. However, the clinical significance of submillimetric differences remains uncertain, as most deviations fall well within the therapeutic tolerance of DBS targeting.
CLINICAL OUTCOMES OF R-DBS FOR PD
While targeting precision is essential, the ultimate measure of DBS success lies in its clinical efficacy. Evidence to date indicates that robot-assisted implantation achieves motor outcomes equivalent to conventional frame-based methods. Both motor function and medication requirements improve after R-DBS, with sustained benefit in activities of daily living and quality of life.
Motor improvement (Unified Parkinson’s disease rating scale [UPDRS]-III)
Across published series summarized in Table 2, R-DBS consistently results in significant postoperative improvements in UPDRS-III scores, typically ranging 20–70% improvement [13,17,18,25,27,28,31,40]. Chang et al. [3] reported that both robotic and conventional frame-based groups demonstrated significant UPDRS-III improvement at 6 months and 1 year (p=0.001 overall), with no significant difference between groups (interaction p=0.119). Paff et al. [27] and Ribault et al. [31] similarly reported comparable motor improvements between R-DBS and conventional DBS groups, with no significant difference in effect sizes. In a large retrospective study of 152 patients, Moran et al. [25] found that “asleep” R-DBS improved mean UPDRS-III scores from 39 to 20.5 (a 47% reduction) at 1 year. Multiple studies further reported that robot-assisted “asleep” and “awake” procedures yielded comparable symptom control, supporting the flexibility of robotic approaches in different perioperative settings [3,13,25]. A systematic review and meta-analysis including 15 studies with 732 PD patients showed improvements in the UPDRS scores [12]. The improvement in UPDRS, UPDRS III, and UPDRS IV was 27.36 (95% CI, 8.57 to 46.15), 14.09 (95% CI, 4.67 to 23.52), and 3.54 (95% CI, -2.35 to 9.43), respectively.
Medication reduction
Levodopa equivalent daily dose (LEDD) reduction is a key secondary outcome, reflecting decreased dopaminergic therapy burden. Paff et al. [27] reported 50% reduction in LEDD at 24 months, while Chang et al. [3] observed a 43% decrease at 12 months, both reaching statistical significance. Several studies [3,4,27] found no significance difference in LEDD reduction between robotic and conventional DBS cohorts. In a meta-analysis, Huang et al. [12] reported a mean postoperative LEDD decrease of 422.31 mg (95% CI, 68.69 to 775.94) after DBS.
Non-motor and quality-of-life measures
Several studies have examined non-motor outcomes and quality-of-life indices such as the PD Questionnaire-39 (PDQ-39). Chang et al. [3] reported significant PDQ-39 improvement in the robot group at 6 months and 1 year (p=0.001 and p=0.032, respectively). Difference between R-DBS and conventional DBS was not-significant (p=0.660).
Moran et al. [25] likewise reported a significant mean PDQ-39 summary improvement of 7.1±14 points, with five of eight subscores showing statistical gains : mobility (+16%), activities of daily living (+30%), emotional well-being (+24%), bodily discomfort (+24%), and stigma (+40%). Additionally, UPDRS II scores improved from 18.9±6 to 10.5±6 (a 44% reduction, p<0.001), while UPDRS IV scores decreased from 7.1±3 to 3.6±3 (a 49% reduction, p<0.001).
Despite these encouraging results, comprehensive assessments of neuropsychological outcomes such as cognition, mood, sleep, and autonomic function remain scarce in the RDBS literature and represent an important area for future investigation.
SAFETY OF R-DBS
Hemorrhage, infection, and device issues
Meta-analyses have consistently shown that R-DBS is not associated with higher complication rates compared with conventional frame-based techniques. Wan et al. [37] found no statistically significant difference in overall complication rates between R-DBS and conventional DBS (OR, 1.79; 95% CI, 0.79–4.05; p=0.17), and hemorrhage rates likewise showed no difference (OR, 0.80; 95% CI, 0.23–2.74; p=0.72).
The meta-analysis by Huang et al. [12] reported an overall complication rate of approximately 12% (95% CI, 3–24%), encompassing hemorrhage, infection, and device-related problems such as lead migration or extension-cable malfunction—figures well within the expected range of conventional DBS [12].
Mechanical or robot-specific failures—including arm collision, trajectory deviation, or registration misalignment—have been reported only sporadically and do not appear to contribute meaningfully to overall surgical risk.
Anesthesia- and MER-related risks
The transition to “asleep” DBS eliminates patient discomfort inherent to awake procedures but introduces anesthesia-related considerations and limits intraoperative neurological testing. Concerns have been raised that general anesthesia might degrade MER signal quality; however, Chang et al. [3] demonstrated reliable MER acquisition under remimazolam-based anesthesia, showing minimal compromise in signal fidelity.
In large comparative series, Jin et al. [13] reported mean radial errors of 0.71±0.25 mm under general anesthesia versus 0.76±0.23 mm under local anesthesia, indicating that asleep DBS maintains acceptable targeting accuracy when intraoperative verification steps are incorporated.
Although multiple MER passes may theoretically increase hemorrhagic risk, evidence from R-DBS cohorts does not suggest any clinically meaningful rise in complications.
OPERATIVE EFFICIENCY AND WORKFLOW
Workflow standardization and advantages
One of the most frequently cited advantages of robotic systems is their contribution to workflow standardization and procedural efficiency. Robotic systems enable preplanned trajectories to be executed with minimal manual adjustment, reduce manual coordinate transformation errors, and allow repeated precise motion [32]. These features reduce intraoperative repositioning and may improve consistency across bilateral implants. Integration with intraoperative imaging further enables trajectory validation or correction before final lead insertion [32].
Giridharan et al. [10] demonstrated that ROSA-based workflows can shorten procedural steps while maintaining accuracy. VanSickle et al. [35] demonstrated that ‘asleep’ R-DBS with intraoperative CT allowed for immediate confirmation of lead placement and reduced operative time compared with conventional awake cohorts. Defrance et al. [5] described a reproducible workflow using the Neuromate system combined with with Oarm cone-beam CT, emphasizing improved intraoperative verification and procedural consistency.
Comparative studies of “asleep” (general anesthesia) and “awake” (local anesthesia) R-DBS techniques found shorter procedure duration in the asleep group, with comparable accuracy, stimulation parameters, and short-term motor improvement across both methods. Clinical outcomes, including UPDRS improvements, did not differ significantly between approaches [13].
Frameless workflows, available on several robotic platforms, enhance patient comfort and reduce time associated with frame fixation, although they may introduce susceptibility to registration error or skull-mounted platform instability [29)].
Procedure time
In a meta-analysis by Wan et al. [37], R-DBS tended to shorten operative or total anesthesia time compared with non-robotic techniques, although heterogeneity across studies was relatively high. Chang et al. [3] reported that robotic set up adds a slight time overhead, however, its enhanced reproducibility and intraoperative correction capability may offset this initial delay.
A clear learning-curve effect has been documented in several centers. Neudorfer et al. [26] reported a reduction in mean operative time from 352.5±49.8 minutes in early cases to 280.5±59.2 minutes in later cases (p<0.001), indicating that procedural familiarity substantially improves efficiency as experience accumulates.
DISCUSSION
Robotic assistance in DBS represents a significant evolution in stereotactic neurosurgery, offering accurate and reproducible trajectory alignment, integrated imaging support, and minimized human error. Current evidence indicates that R-DBS achieves targeting accuracy comparable or slightly superior to conventional frame-based techniques while providing workflow advantages and maintaining a favorable safety profile.
While most studies report high targeting accuracy for RDBS, with mean errors clustering around approximately 1 mm, several reports describe higher mean targeting errors in the range of approximately 1.3–1.5 mm. This variability likely reflects a combination of technical, methodological, and procedural factors rather than fundamental limitations of robotic guidance. Differences in robotic system design and generation, image acquisition and registration protocols, frame-based versus frameless workflows, and definitions of targeting error (e.g., radial versus three-dimensional error) may substantially influence reported accuracy. In addition, surgeon experience and institutional learning-curve effects, particularly during early adoption phases, as well as patient-related factors such as brain shift and ventricular size, may further contribute to inter-study differences. These findings underscore that reported targeting accuracy should be interpreted within the context of system-specific workflows and methodological heterogeneity across studies.
A central question, however, is whether marginal improvements in targeting precision – often fractions of a millimeter – translate into measurable clinical benefit. Current data do not demonstrate clear superiority of R-DBS in terms of motor outcomes compared with traditional DBS. Nevertheless, its improved reproducibility, reduced inter-operator variability, and streamlined procedural workflow may provide indirect advantages, particularly in high-volume centers or institutions adopting “asleep” or frameless paradigms.
Despite the current parity between robot-assisted and conventional DBS in terms of targeting accuracy and clinical outcomes, there remain several reasons why continued development and evaluation of robotic DBS may be justified. Robotic systems provide a framework for procedural standardization that may enhance consistency across surgeons and institutions which is an aspect not fully captured by mean accuracy or outcome metrics alone. As DBS programs expand and surgical workflows evolve, reproducibility and workflow reliability may become increasingly important determinants of overall treatment quality. Importantly, the absence of demonstrated superiority should not be interpreted as technological futility, but rather as an indication that current robotic platforms have yet to overcome specific technical and workflow-related constraints that may be necessary to achieve clearer clinical advantages.
Building on these considerations, each currently available robotic system carries inherent technical limitations. Intraoperative imaging integration is not uniformly supported across platforms, and responding to intraoperative contingencies may be less flexible than with conventional frame-based approaches. Dedicated technical personnel and system calibration requirements may further limit flexibility. Addressing these constraints through improved imaging integration, greater intraoperative adaptability, and enhanced workflow versatility will be essential for robotic DBS to demonstrate more definitive advantages.
PD-specific implications of R-DBS
PD presents several disease-specific characteristics that make robotic assistance particularly relevant in DBS surgery. Patients with PD are typically older, often have advanced motor symptoms, and may exhibit anxiety, discomfort, or limited cooperation during prolonged awake procedures. In addition, bilateral implantation is commonly required, increasing procedural complexity and patient burden when traditional awake, frame-based techniques are used.
In this context, R-DBS facilitates image-guided targeting with high reproducibility and enables reliable asleep DBS workflows that may be especially advantageous for PD patients. While MER remains an important physiological confirmation tool, its reliability can be influenced by patient cooperation and intraoperative conditions. Robotic systems, combined with high-quality preoperative imaging and intraoperative image verification, allow accurate anatomical targeting that may reduce reliance on prolonged awake testing in selected PD patients.
Furthermore, unlike other movement disorders where awake intraoperative testing may be critical for immediate symptom feedback, the necessity of awake surgery in PD is comparatively lower, as motor outcomes are typically assessed postoperatively through programming and medication adjustment. As a result, asleep and image-guided robotic DBS workflows may offer a favorable balance between accuracy, patient comfort, and procedural efficiency in PD.
Taken together, these considerations suggest that the relevance of robotic DBS in PD lies not merely in prevalence, but in the alignment between disease-specific clinical characteristics and the technical strengths of reproducible, image-guided stereotactic surgery.
Limitations & future advancements
Several limitations in the current literature hinder comprehensive interpretation. Long-term follow-up data remain limited, and most existing studies are retrospective and single-center, employing heterogeneous robotic systems, targeting protocols, and outcome measures that impede direct comparison. Furthermore, practical considerations including the learning curve, equipment cost, maintenance burden, and the need for seamless integration into existing operating-room infrastructure continue to influence broader adoption.
Future advancement will rely on prospective, multicenter randomized studies and standardized reporting frameworks that enable rigorous cross-platform comparison. Continued technological innovation is expected to focus on real-time intraoperative feedback, including imaging and electrophysiologic confirmation, as well as AI-assisted trajectory planning and closed-loop DBS paradigms capable of dynamically adapting stimulation parameters based on neurophysiologic feedback.
Overall, R-DBS continues to show promise as a safe, accurate, and efficient alternative to conventional stereotactic techniques. However, long-term comparative studies remain essential to determine whether these technical refinements can translate into superior clinical outcomes. With ongoing innovation and refinement, R-DBS may have the potential to eventually evolve to replace traditional frame-based methods
CONCLUSION
R-DBS for PD achieves targeting accuracy comparable to conventional stereotactic techniques, with equivalent clinical efficacy, non-inferior safety, and potential advantages in workflow standardization and reproducibility. Although evidence of superiority in clinical outcomes remain inconclusive, robotic DBS offers increasing value through improved procedural consistency and ongoing technological innovation, supporting its growing role in advanced DBS centers. Continued technological refinement and high-quality comparative studies will be essential to determine whether these strengths can translate into clear and durable clinical advantages
Notes
Conflicts of interest
J.I.L. is the recipient of a research grant from Koh Young Technology, but Koh Young had no role in the preparation, interpretation, or conclusions of this review. This article does not imply or assert any superiority among existing robotic systems. Artificial intelligence assistance was used solely for English language editing and refinement during manuscript preparation. The authors take full responsibility for the content and accuracy of the final version. No potential conflict of interest relevant to this article was reported.
Informed consent
This type of study does not require informed consent.
Author contributions
Conceptualization : JIL; Data curation : KWC; Formal analysis : KWC; Methodology : KWC; Project administration : JIL; Visualization : KWC; Writing - original draft : KWC; Writing - review & editing : KWC, JIL
Data sharing
None
Preprint
None
