An Efficient Method for Aneurysm Volume Quantification Applicable in Any Shape and Modalities
Article information
Abstract
Objective
Aneurysm volume quantification (AVQ) using the equation of ellipsoid volume is widely used although it is inaccurate. Furthermore, AVQ with 3-dimensional (3D) rendered data has limitations in general use. A novel universal method for AVQ is introduced for any diagnostic modality and application to any shape of aneurysms.
Methods
Relevant AVQ studies published from January 1997 to June 2019 were identified to determine common methods of AVQ. The basic idea is to eliminate the normal artery volume from 3D model with the aneurysm. After Digital Imaging and Communications in Medicine (DICOM) data is converted and exported to stereolithography (STL) file format, the 3D STL model is modified to remove the aneurysm and the volume difference between the 3D model with/without the aneurysm is defined as the aneurysm volume. Fifty randomly selected aneurysms from DICOM database were used to validate the different AVQ methods.
Results
We reviewed and categorized AVQ methods in 121 studies. Approximately 60% used the ellipsoid method, while 24% used the 3D model. For 50 randomly selected aneurysms, volumes were measured using 3D Slicer, RadiAnt, and ellipsoid method. Using 3D Slicer as the reference, the ratios of mean difference to mean volume obtained by RadiAnt and ellipsoid method were -1.21±7.46% and 4.04±30.54%, respectively. The deviations between RadiAnt and 3D Slicer were small despite of aneurysm shapes, but those of ellipsoid method and 3D Slicer were large.
Conclusion
In spite of inaccuracy, ellipsoid method is still mostly used. We propose a novel universal method for AVQ that is valid, low cost, and easy to use.
INTRODUCTION
The coil volume embolization ratio (VER) was first introduced in 1997 by Satoh et al. [27]. The authors considered an aneurysm as an ellipsoid and used the mathematical equation for ellipsoid volume to calculate aneurysm volume. However, this method of aneurysm volume quantification (AVQ) may contain errors and is not compatible for use with irregularly shaped or fusiform aneurysms [26,32]. Nonetheless, it has been widely used in studies and applied to commercial software for calculating coil packing density. Other AVQ methods have been introduced and improved upon with 3-dimensional (3D) angiographic data, but these also have limitations related to imaging modality, angiography equipment (workstations), and the particular software utilized [2,5,6,8,9,23,26]. Therefore, the applicability of these methods to clinical use and research is also somewhat limited. In order to overcome these problems, we introduce a novel universal method for AVQ that can be used with any diagnostic modality and is applicable to any aneurysm shape.
MATERIALS AND METHODS
This retrospective study was approved by the Institutional Review Board of Dankook University Hospital (2020-07-017) before data collection began. Because of the retrospective design of the study the patient consent for this study was waived by the Institutional Review Board.
Literature review
The PubMed database was searched using the keywords “VER”, “packing density”, and “aneurysm volume” to identify relevant AVQ studies published between January 1997 and June 2019. Each study was reviewed and categorized according to AVQ method and year of publication. When the method was not clearly described, we followed the reference cited in the manuscript’s definition of aneurysm volume. Based on our review, we established the following AVQ categories : 2-dimensional (2D) assessment, ellipsoid method, AngioCalc and AngioSuite, Workstation, commercial software, and phantom. The 2D assessment category includes non-quantitative methods that use plain radiographic images such as the Raymond-Roy occlusion classification [30]. The ellipsoid method uses the mathematical equation for ellipsoid volume. The AngioCalc and Angiosuite category is comprised of the web-based and mobile device applications AngioCalc and AngioSuite Neuro Edition, respectively. In brief, they each use the ellipsoid method and automatically calculate coil volume and VER. Workstation is software provided for angiography suite equipment. The commercial software category is comprised of various types of mathematical, computer-aided design (CAD), and computational fluid dynamics (CFD) software. The phantom category is comprised of experimental methods that use phantom models for AVQ.
Definition of aneurysm volume
Aneurysm volume is usually defined as the volume of the outpouching abnormal vascular lesion above the aneurysm neck as in Fig. 1A. However, determination of the aneurysm neck is rather complicated in fusiform aneurysms and application of the same volume definition to both saccular and fusiform aneurysms may not apply (Fig. 1B). For this study, we defined aneurysm volume as the remnant volume after removing the normal arterial component, which can be applied to aneurysms of any shape or type (Fig. 1C and D). The data of 50 randomly selected aneurysms from our 3D digital subtraction angiography (DSA) Digital Imaging and Communications in Medicine (DICOM) database which were collected from March 2018 to June 2019 were used to compare and validate the different AVQ methods.
Aneurysm volume measurement
The basic idea of our novel universal 3D AVQ method is to differentiate the normal artery volume from the aneurysm prior to volume measurement (Fig. 2). Three steps are involved : data conversion, 3D model modification, and aneurysm volume measurement (Supplementary Video 1).
Conversion of DICOM data to a 3D model and model post-processing
Open-source 3D Slicer software, version 4.10.2 (http://www.slicer.org) and RadiAnt DICOM viewer software, version 5.0.0 (http://www.radiantviewer.com; Medixant, Poznań, Poland) were used to convert DICOM data into a 3D model [10]. We then compared the 3D Slicer, RadiAnt, and ellipsoid AVQ methods using the 50 aneurysms from our DICOM database. In 3D Slicer, the vasculature was isolated by modifying the threshold tool. The threshold level was adjusted until the major cerebral blood vessels showed clear and distinct margins on axial, sagittal, and coronal images. After rendering the vascular 3D model, it was exported in stereolithography (STL) file format. In RadiAnt, the 3D model was generated with the 3D volume rendering tool, adjusted by controlling the window level, and also exported in STL file format. The segmentation of 3D models can be done in current step using 3D slicer and RadiAnt. But it can also be done in the next step, which is 3D STL model modification.
3D STL model modification
Once STL files were exported, the 3D STL model was opened with Meshmixer software, version 3.5 (http://www.meshmixer.com; Autodesk, Inc., San Rafael, CA, USA). The aneurysm was digitally removed and converted to a normal arterial vascular feature by applying the edit and sculpt tool. During the modification of a STL file, it is important not to make any modification of any other vessel components. Two 3D models of each patient were generated, one which included the aneurysm and one which excluded it and contained digitally reconstructed normal vasculature in its place.
Aneurysm volume measurement
The volumes of the two 3D models were measured with open-source Meshlab software, version 2016 (Visual Computing Lab, ISTI - CNR, Pisa, Italy). The difference between the two volumes was defined as the aneurysm volume.
Aneurysm morphology parameters and the classical ellipsoid method for AVQ
Aneurysm height, neck size, and aspect ratio were measured as described by Dhar et al. [7]. The classical aneurysm volume measurement method that considers the aneurysm as an ellipsoid (ellipsoid method) was followed as described by Satoh et al. [27].
Statistical analysis
Statistical analysis was performed with SPSS software, version 25 (IBM, Inc., Armonk, NY, USA). Descriptive statistics of aneurysmal morphologic parameters are presented as means±standard deviation (SD). The agreement between AVQ methods was assessed with Bland-Altman plots using a 10% difference limit of acceptability with 3D Slicer data as the reference value [3,6]. Difference ratios in the Bland-Altman plots were converted to percentile values. AVQ methods were compared using the paired t-test. p<0.05 was considered significant. As the ellipsoid method is not applicable to fusiform aneurysms, data of four cases of fusiform aneurysms were discarded in comparing the intermethod agreement through the Bland-Altman plot.
RESULTS
The PubMed literature search identified 215 studies published between January 1997 and June 2019; 95 were not relevant for the purposes of this investigation and therefore excluded. Finally, 120 studies were reviewed and classified. Although AngioCalc and AngioSuite provide an interface by which users can manually input aneurysm volume, they calculate AVQ based on the ellipsoid method, as do several other AVQ methods. Hence, approximately 60% of all studies included in this investigation used the ellipsoid method; additionally, approximately 9% of the studies used 2D assessment and 25% used a 3D model, which does not exceed one third on a 5-year basis (Fig. 3 and Supplementary Data 1).
Volume of the 50 aneurysms randomly selected from our DICOM database were measured using 3D Slicer, Radiant DICOM viewer, and the ellipsoid method. Forty-six sacular aneurysms were included in comparison analysis of intermethod agreement; 16 of these were lobulated. The remaining four aneurysms were fusiform. Mean maximal aneurysm diameter and diameter of the aneurysm neck were approximately 6 mm and 4 mm, respectively. Using 3D Slicer as a reference, the ratio of mean difference to mean volume obtained using RadiAnt and the ellipsoid method were -1.21±7.46% and 4.04± 30.54%, respectively. Although the methods were used with aneurysms of many different shapes and configurations, the SDs in RadiAnt and 3D Slicer were very close. When comparing the ellipsoid method to 3D Slicer, the SD difference between methods increased in lobulated and fusiform aneurysms compared to non-lobulated aneurysms. The ratio of volume difference less than 10% between RadiAnt to 3D Slicer is almost twice of Ellipsoid method to 3D Slicer (Table 1). This implies that intermethod agreement is higher between RadiAnt and 3D Slicer than the ellipsoid method and 3D Slicer.
The Bland-Altman plots demonstrate the degree of agreement among the AVQ methods using 3D Slicer as the reference. The RadiAnt-3D Slicer plot shows that most mean volume differences were under 10%. The difference in SD between RadiAnt and 3D Slicer was less than the difference between the ellipsoid method and 3D Slicer (Fig. 4). Note that the SD of RadiAnt and 3D Slicer is within 10% of mean volume difference while the SD of the ellipsoid method and 3D Slicer is beyond 10% of mean volume difference. In the subanalysis based on aneurysm shape, the SD difference between RadiAnt and 3D Slicer was small. However, when comparing the difference between the ellipsoid method and 3D Slicer, SD tended to be larger in lobulated and fusiform aneurysms (Fig. 5). The SD of RadiAnt and 3D Slicer is within 10% of mean volume difference while the SD of the ellipsoid method and 3D Slicer is beyond 10% of mean volume difference in both saccular and irregular shape of the aneurysm. Additional Bland-Altman plot of four cases of fusiform aneurysms is also demonstrated in Fig. 5.
DISCUSSION
The concept of VER was introduced to represent the volume of embolized material (e.g., coils) versus aneurysm volume. Although the term “packing density” is also widely used to represent the same concept, it refers to radiographic opacity inside an aneurysm visualized on 2D plain angiography. The widely used Raymond-Roy classification is based upon this concept [21,24,29,34]. It is relatively easy to calculate coil volume by assuming the coil as a cylinder. Traditionally, AVQ has been calculated by considering the aneurysm as an ellipsoid [27]. However, the accuracy of this method has been questioned and debated [26,32]. Many authors have pointed out that the ellipsoid method may underestimate or miscalculate the volume, particularly in wide-neck and lobulated aneurysms. Nevertheless, it has been widely used due to its simplicity. Low VER and remnant volume have been associated with aneurysm recurrence in previous studies [16,17,22,25,28,33]. Although many studies have been mentioning VER, the majority still use the ellipsoid method for AVQ (Fig. 3 and Supplementary Data 1). To achieve better accuracy, a method using 3D rendered data for AVQ has been recommended.
There are several issues to point out regarding determination of aneurysm volume. First, aneurysm volume can only be measured indirectly; it cannot be measured directly in vivo. Therefore, DICOM data must be used to determine aneurysm volume. However, different imaging modalities and patient condition (e.g., blood pressure, pulse, etc.) may affect the indirect measurement of aneurysm volume. Second, aneurysm volume is usually defined as the volume of the abnormal outpouching vascular lesion above the aneurysm neck. However, in fusiform aneurysms, determination of the aneurysm neck is rather complicated, and applying the same definition of volume to both saccular and fusiform aneurysms is rather difficult. Therefore, we elected to define aneurysm volume as the remnant volume after removing the normal arterial volume. This definition can be applied to aneurysms of any shape.
A universal method for AVQ should fulfill several conditions. First, it should have a low barrier to entry for general use; the method should be easy to apply and interpret and cost little. Second, it should be able to be performed expeditiously; if it requires >30 minutes to calculate, the method may not be accepted for widespread use. Third, it should be able to be used with various imaging modalities, such as computed tomography angiography (CTA), magnetic resonance angiography (MRA), and DSA. Finally, the method should be accurate and reproducible and able to be used for any aneurysm shape or type.
For the universal AVQ method proposed here, the first step is to convert DICOM data into a 3D model. Since 3D Slicer has been commonly used in various medical and bioengineering fields, we elected to use it as the reference method [1,11,12,20,31]. Although 3D Slicer has been previously used to define cerebral vascular anatomy and obtain vascular parameters, to our knowledge, it has not been used to measure aneurysm volume [4,13-15,18,19]. After the 3D model is generated in 3D Slicer, it is possible to measure the aneurysm volume by segmentation. The segmentation process in 3D Slicer is easily implemented for saccular aneurysms, but not for fusiform aneurysms or aneurysms of a curved vessel. Therefore, we exported the 3D model in STL file format and sculped it with Meshmixer, software which was designed for 3D printing of STL files. Since Meshmixer provides various tools to edit 3D STL files, it can be used for aneurysms of any shape. However, volume cannot be measured with Meshmixer; this was accomplished using Meshlab. Although three separate programs are required, all are free or open-source software available on the internet.
During the conduction of this study, we found a considerable learning curve with using 3D Slicer : a 3D STL model could be generated in <3 minutes by experienced users but beginners required almost 40. However, RadiAnt can also export 3D rendered data in STL file format. Its interface is very intuitive and easy to learn and it requires fewer processing steps: experienced users and beginners alike required <5 minutes to generate a 3D STL model. We validated high intermethod agreement between 3D Slicer and RadiAnt. In addition, both 3D Slicer and RadiAnt support various imaging modalities including CTA, MRA, and 3D DSA. Pros and cons of each software and AVQs are presented in Table 2.
The Bland-Altman plot shows the SD of volume difference in AVQ methods performed with 3D STL models in within 10% of mean volume difference which represent a high degree of intermethod agreement (Figs. 4 and 5). The specific software used to convert DICOM data into a 3D STL model does not appear to affect the result, even in aneurysms of various shapes, because the DICOM data itself functions as a representative depiction of the vasculature. However, the ellipsoid method showed a lower degree of agreement than 3D STL which the SD is beyond 10% of mean volume difference which represents the result of low intermethod agreement (Figs. 4 and 5). The intermethod agreement was even worse in multi-lobulated and fusiform aneurysms (Fig. 5). This indicates that AVQ with a 3D STL model is more reliable in any type of aneurysm than the ellipsoid method. Therefore, AVQ using a 3D STL model has the potential to become a universal method in future studies investigating the correlation between aneurysm recurrence and VER.
We encountered a similar AVQ idea in a study by Costalat et al. [6]. The authors used software that automatically calculates the erased volume from a 3D rendered model: the aneurysm is erased until a normal artery form is reached. Although this resembles the basic concept of previous studies, their study was the only one using this method according to our PubMed search. The reproducibility and accuracy of the method in Costalat et al. [6] were high, but its use would be limited as it requires special proprietary software. Although most angiography suite workstations provide various methods of AVQ, if the aneurysm is fusiform or contains multiple lobes, it may not be accurately detected. In addition, CAD and CFD software have also been used for AVQ, but their high cost and level of skill required heighten the entry barrier for widespread general use (Table 2).
An additional advantage of using our proposed novel AVQ method is derived from the data conversion process which exports vasculature data into a 3D STL model. This STL file format was developed by 3D Systems (Rock Hill, SC, USA) to store 3D geographic data in the 1980s and has been used in various fields including CAD and 3D printing. Therefore, a 3D STL model can be exported to CFD software to investigate intravascular fluid dynamics and to 3D printing software to generate phantom models. Use of this novel AVQ method may lead to improved understanding of aneurysmal fluid dynamics and inspire new clinical and scientific advancement.
Recently, the interest in endovascular treatment of intracranial aneurysm has shifted to flow diverters. For this reason, the interest of VER in endovascular coiling seems to be decreased. However, aneurysm coiling is still the mainstream in endovascular treatment and VER is a crucial factor in recurrence of coiled aneurysm. For objective calculation of VER, reasonable method for AVQ is essential. Furthermore, to understand the nature of the aneurysm and endovascular coiling, quantification and interpretation of the aneurysm and surrounding vascular anatomy is crucial, especially in the era of CFD. Therefore, we assure that refining the definition of the aneurysm and developing an AVQ for various shapes of aneurysms can give us more information in understanding the nature of intracranial aneurysms.
CONCLUSION
Even in the era of 3D angiography, most clinical and scientific studies still rely on the ellipsoid method for AVQ despite its problems with accuracy and incompatibility for use with fusiform and irregularly shaped aneurysms. To overcome these problems, we propose a novel universal method for AVQ that utilizes user-friendly open source 3D STL model editing software that is available on the internet. We believe this novel method will be helpful in clinical practice and in future scientific studies.
Notes
CONFLICTS OF INTEREST
No potential conflict of interest relevant to this article was reported.
INFORMED CONSENT
This type of study does not require informed consent.
AUTHOR CONTRIBUTIONS
Conceptualization : JC
Data curation : JC
Formal analysis : JC
Methodology : JC
Project administration : JC
Visualization : JC, JHK
Writing - original draft : JC, JHK
Writing - review & editing : JC, JHK