| Home | E-Submission | Sitemap | Editorial Office |  
top_img
Journal of Korean Neurosurgical Society 2010;47(2): 128-133.
doi: https://doi.org/10.3340/jkns.2010.47.2.128
The Variable Ellipsoid Modeling Technique as a Verification Method for the Treatment Planning System of Gamma Knife Radiosurgery.
Beong Ik Hur, Byung Kwan Choi, Soon Ki Sung, Won Ho Cho, Seung Heon Cha, Chang Hwa Choi
Department of Neurosurgery & Medical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Korea. spine@pusan.ac.kr
ABSTRACT
OBJECTIVE
The secondary verification of Leksell Gamma Knife treatment planning system (LGP) (which is the primary verification system) is extremely important in order to minimize the risk of treatment errors. Although prior methods have been developed to verify maximum dose and treatment time, none have studied maximum dose coordinates and treatment volume.
METHODS
We simulated the skull shape as an ellipsoid with its center at the junction between the mammillary bodies and the brain stem. The radiation depths of the beamlets emitted from 201 collimators were calculated based on the relationship between this ellipsoid and a single beamlet expressed as a straight line. A computer program was coded to execute the algorithm. A database system was adopted to log the doses for 31x31x31 or 29,791 matrix points allowing for future queries to be made of the matrix of interest.
RESULTS
When we compared the parameters in seven patients, all parameters showed good correlation. The number of matrix points with a dose higher than 30% of the maximal dose was within +/- 2% of LGP. The 50% dose volume, which is generally the target volume, differs maximally by 4.2%. The difference of the maximal dose ranges from 0.7% to 7%.
CONCLUSION
Based on the results, the variable ellipsoid modeling technique or variable ellipsoid modeling technique (VEMT) can be a useful and independent tool to verify the important parameters of LGP and make up for LGP.
Key Words: Gamma knife radiosurgery; Treatment planning system; Quality assurance
TOOLS
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
4
Crossref
4
Scopus
2,133
View
23
Download
Related article
Editorial Office
1F, 18, Heolleung-ro 569-gil, Gangnam-gu, Seoul, Republic of Korea
TEL: +82-2-525-7552   FAX: +82-2-525-7554   E-mail: office@jkns.or.kr
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © Korean Neurosurgical Society.                 Developed in M2PI
Close layer